Specifications | Refrigerated Blood Bank Centrifuge | Refrigerated Blood Bank Centrifuge | ||
Model No | mLabs-RB01 | mLabs-RB02 | mLabs-RB03 | mLabs-RB04 |
Capacities | 6 x 550 mL
Blood bags |
8x 550 mL
blood bags |
12 x 500 mL
Blood Bags |
16 x 500 mL
blood bags |
Maximum speed | 5,000 rpm | 4,700 rpm | ||
Maximum RCF | 7,295 x g | 7,187 x g | ||
Drive system | High torque brushless drive technology | High torque brushless drive technology | ||
Imbalance tolerance | 125 g in opposite loads | 125 g in opposite loads | ||
Accumulated centrifugal effect (ACE) integrator | Yes | Yes | ||
Refrigeration off when door opens;
wind shielded rotors |
Refrigeration off when door opens;
wind shielded rotors |
|||
Green technology | ||||
Control | 7 inch touch screen interface | 7 inch Touch touch screen interface | ||
Accel/decel profiles and existing protocol emulation | 11 accel, 12 decel, multiple emulation profiles | 11 accel, 12 decel, multiple emulation profiles | ||
Modes | At start, at speed, time start | At start, at speed, time start | ||
Step-runs | Yes, 30 profile/speed/time triplets, up to 3 steps each | Yes, 30 profile/speed/time triplets, up to 3 steps each | ||
Maintenance tracking log | Yes | Yes | ||
Protocol traceability | Yes, built-in Centri-Track run logging | Yes, built-in Centri-Track run logging | ||
Performance range | Speed 300–5,000 rpm; RCF 26-7,295 x g | Speed 300–4,700 rpm; RCF 29-7,187 x g | ||
Run time | 99 hours 59 min 59 sec (1 second increment) | 99 hours 59 min 59 sec (1 second increment) | ||
Pre-cooling function | Yes | Yes | ||
Temperature set range | –20 to 40°C, adjustable to 1°C | –20 to 40°C, adjustable to 1°C | ||
Other functions |
Multilingual selection, on-board training videos, user logging, user lock-out, automatic door opening, automatic rotor ID, on screen-display for imbalance, over temperature, stainless steel chamber, guidance display for
error messages |
|||
Dimensions (H x W x D) | 1015 x 816 x 900 mm (39.9 x 32.1 x 35.4 in) | 1015 x 816 x 900 mm (39.9 x 32.1 x 35.4 in) | ||
Weight
|
230V: 475 kg/1047.2 lbs; | |||
230V: 475 kg/1047.2 lbs; | 230V Air Cooled: 479 kg/1056 lbs; | |||
400V: 495 kg/1091.3 lbs | 400V: 495 kg/1091.3 lbs | |||
Heat output | < 2 kW | < 2 kW | ||
Power consumption | < 5,400 VA | < 5,400 VA | ||
Certifications | ISO 9001, ISO13485, WHO GMP, CE and FDA | |||
Standards |
EN 61010-1 3rd Edition, IEC 61010-2-020 3rd Edition, IEC 61010-2-101 3rd Edition, EN 61326-1 Class B, EN 61326-2-6, EN 62304, EN 62366, EN ISO 14971, EN ISO 13485 | |||
Warranty |
2 yrs unit, 5 yrs power-train (motor shaft and drive), 5 yrs refrigeration | 2 yrs unit, 5 yrs power-train (motor shaft and drive), 5 yrs refrigeration |
Blood Bank Refrigerated Centrifuge 6 and 12 Bags
In Stock
Highlights:
BRAND – mLabs
- 4800 R P M with 10 RPM increment
- Continues read out of working RCF Value
- Temperature control with Working temperature in the range of -5 °C to 25 °C depending on the speed of rotations
- 99 program memory, brush less induction motor drives
- 1 to 120 minutes timer programmable in 1 minute increment with infinite mode
- User programmable Acceleration and Deceleration time range from 40 seconds to 450 seconds
- Mechanical and Electronic lid lock with indicator,
- Imbalance indicator
- LCD display of set and working parameters.
- Specially designed cooling system allow pre-cooling of rotor head and also allow the cooling system to work even when the centrifuging operation is complete, thereby eliminating damage to heat sensitive material.
- The Centrifuge chamber is made of stainless steel and is effectively insulated and separated from other parts
- Counter Balanced & fall safe Lid with hydraulic system.
- Max. Speed 4800 RPM , Max RCF. 6000x g depending on rotor heads. (without rotor head and carriers).
- 6 Place Swing out rotor suitable for carrying 6 Nos. metal buckets Maximum speed 4500 RPM . Centrifugal force 5000-6000g
Description